Short-read sequencing
Sequencing by synthesis with reversible terminators, from benchtop instruments (millions of reads) to production instruments (billions).
Our ruleSimulations prepare you for the bench but don't replace hands-on practice.
Illumina
Sequencing by synthesis with reversible terminators, from benchtop instruments (millions of reads) to production instruments (billions).
1Sequencing by synthesis
Library molecules carry adapters that let them hybridise to oligos on a glass flow cell. Each molecule is copied in place into a cluster of ~1000 identical copies (bridge or exclusion amplification).
Cluster density
Too few clusters waste the flow cell; too many overlap and can't be told apart.
Base quality
Q = −10·log₁₀(error). Q30 means one error in 1000 bases. Runs are judged on % of bases ≥ Q30.
Passing filter
Clusters with clean enough signal in the first cycles to be kept.
2What the instrument needs to work
- An accurately quantified, denatured library loaded at the right concentration (qPCR counts only molecules that can form clusters).
- Base diversity: in every cycle the instrument expects all four colours across clusters. Amplicon libraries that start with the same primer sequence need PhiX spiked in (10–20%) to supply diversity.
- No adapter dimers: short fragments cluster efficiently and steal reads.
3Single-end vs paired-end
Single-end runs read one end of each fragment. Paired-end runs turn the molecules around and read the other end too, which doubles the data per cluster, improves mapping and lets you detect insertions, deletions and rearrangements.
Key terms
Practise these as flashcards →DNA fragments with sequencing adapters and indexes attached.
A short sequence identifying the sample, so many libraries can be pooled and separated afterwards (demultiplexing).
A clonal group of copies of one library molecule on the flow cell.
Log-scaled probability that a base call is wrong: Q20 = 1%, Q30 = 0.1%.
A well-characterised viral library spiked in as a control and to add base diversity.
Text format holding each read's sequence and per-base quality scores.
When to use it
- Almost every genomics assay: whole genomes, exomes, RNA-seq, ChIP-seq, ATAC-seq, amplicon and single-cell libraries.
- When you need many reads at low cost per base and reads of 50–300 bases are enough.
- When accuracy per base matters (short reads are ~99.9% accurate).
Compared with alternatives
Reads of 10–100 kb resolve repeats and structural variants; lower throughput per run.
One ~800 bp read per reaction; ideal for checking a plasmid or PCR product.
How the libraries are made.
Checking size and molarity before loading.
What happens to the FASTQ files afterwards.
Sub-techniques
Choosing one here switches the whole page, including the protocol and the lab, to it.
Sequencing by synthesis with reversible terminators, from benchtop instruments (millions of reads) to production instruments (billions).