SEC-MALS

Determine absolute molecular weight and oligomeric state.

molar massSimulator readyCore simulatorEngine: Chromatography, purification & mass spec

Our ruleSimulations prepare you for the bench but don't replace hands-on practice.

1Weighing molecules as they leave a column

Size-exclusion chromatography separates molecules by how large they are in solution, not by how heavy. A column calibrated with globular standards turns an elution volume into a mass only if your protein has the same shape and density as those standards; an elongated monomer, a flexible protein or a detergent-solubilised membrane protein can elute where something twice its mass would.

2Reading the result

Mass across the peak
Plotted over the chromatogram. A pure species gives a flat line; a mass that falls from the front to the back of a peak is co-eluting species or an equilibrium mixture; a mass that tilts the same way across every peak is misaligned detectors.
Weight-average mass (Mw)
The mass of a peak computed from the summed signals across it. Mw/Mn close to 1.00 means a monodisperse peak.
dn/dc
How much the refractive index rises per unit of concentration: about 0.185 mL/g for unmodified proteins, lower for glycans, lipids and detergents, which is why conjugates need a protein-conjugate analysis with UV and RI together.
Delay volume and band broadening
Each detector sees the peak a little later and a little wider than the one before. Both are measured with a monodisperse standard (often BSA) and corrected before slices from different detectors are combined.

3How it fails

  • A mass that slopes across a symmetric single-species peak: the inter-detector delay or the band-broadening correction is wrong.
  • A mass consistently too low or too high by a fixed factor: wrong concentration (extinction coefficient, dn/dc) or calibration constant.
  • One peak whose mass falls steadily across it: species co-elute, often because the column's range does not bracket them.
  • Ragged LS baseline, spikes and scattered masses: a column still shedding particles, unfiltered buffer, or a sample that was not spun.
  • Too little protein: light scattering grows with concentration × mass, so small proteins need more material than large ones.
Rayleigh ratio

The normalised intensity of light scattered at an angle; the excess over the solvent is what MALS measures.

Zimm equation

Relates the excess Rayleigh ratio to concentration, molar mass, Rg and the second virial coefficient; the basis of all static light-scattering mass determination.

Radius of gyration (Rg)

From the angular dependence of the scattering, measurable only for particles above roughly 10 nm with visible light.

Protein-conjugate analysis

Using UV and RI together, with each component's extinction coefficient and dn/dc, to split the mass of a glycoprotein or protein–detergent complex into protein and modifier.

Void volume (V₀)

Where everything too large to enter the pores elutes; MALS still weighs what comes out there, so aggregates are quantified rather than lumped together.

Normalisation

Calibrating the detectors at different angles against the 90° detector with an isotropic scatterer, so that they agree for small molecules.

When to use it

  • Deciding the oligomeric state of a protein in solution, especially when its elution volume is ambiguous (elongated, flexible, glycosylated or detergent-solubilised).
  • Measuring the mass and amount of aggregates, fragments and complexes, for example before structural work or as a quality check of a biologic.
  • Following complex formation: injecting partners alone and mixed to see a peak of the expected combined mass.

Compared with alternatives

Chromatography polishing →

Preparative SEC purifies and its calibration gives an apparent size; add MALS when you need to know what the peaks actually weigh.

Analytical ultracentrifugation →

Measures size and heterogeneity without a column matrix, so weak or rapidly exchanging complexes are not pulled apart; slower and needs more expertise.

Dynamic light scattering →

A quick check for aggregates in a cuvette with no separation; gives a hydrodynamic size, not a mass.

Mass spectrometry proteomics →

Gives an exact covalent mass and modifications, but not the native oligomeric state in solution.