1Seeing inside an animal without opening it
In vivo imaging follows the same animal over days or weeks: a tumour's growth, the spread of labelled cells, the remodelling of bone around an implant. Each technique trades depth, resolution and specificity differently.
Bioluminescence
Luciferase cells plus injected luciferin emit light; almost no background, so very sensitive, but blurred and depth-dependent.
Fluorescence
Excite a reporter and collect its emission; works best in the near infrared, where tissue absorbs and autofluoresces least.
Micro-CT
Three-dimensional X-ray images with voxels down to ~10 µm; quantitative for bone, at the price of a radiation dose.
2What the numbers depend on
Bioluminescence depends on substrate kinetics: after an intraperitoneal luciferin injection the signal rises over minutes, peaks and decays, so every mouse must be imaged at the same time after injection, found from a kinetic series. It also falls with depth, dark fur and a cold animal.
3Anaesthesia, welfare and the 3Rs
Animals are imaged under inhalational anaesthesia on a warmed stage with eye lubricant, and recovered in a warmed cage under observation; long sessions need monitoring of breathing and temperature. Imaging needs institutional ethical approval and trained staff like any other procedure.
Key terms
Practise these as flashcards →Photons per second from a region of interest: the exposure-independent unit for bioluminescence.
The rise and fall of the bioluminescent signal after substrate injection; imaging is timed to its plateau or peak.
Pixels at the camera's maximum count, where signal no longer rises with light.
Roughly 650–900 nm, where haemoglobin and water absorb least and tissue is most transparent.
Fluorescence from the tissue or diet itself, such as chlorophyll in chow, that competes with the label.
Blurring of a structure over neighbouring voxels when it is not much larger than the voxel.
Bone volume fraction and trabecular thickness, the standard micro-CT measures of trabecular bone.
When to use it
- Following tumour growth or metastasis in the same animals over time with luciferase-expressing cells.
- Tracking a fluorescently labelled probe or cell population near the surface or in the near infrared.
- Measuring bone architecture or implant integration longitudinally.
Compared with alternatives
Callipers measure subcutaneous tumours directly; imaging adds deep or metastatic disease and implants hidden from view.
For activity of individual cells in a window or a slice, microscope-based imaging gives cellular resolution that whole-body imaging cannot.
How to choose a reporter's spectrum and brightness, which decides what can be seen through tissue.
Sub-techniques
Choosing one here switches the whole page, including the protocol and the lab, to it.
Luciferase-expressing cells imaged in an anaesthetised animal after a luciferin injection, in a light-tight camera such as an IVIS. Highly sensitive with almost no background, but semi-quantitative: light falls steeply with depth and depends on substrate timing.
- Image at a fixed time after luciferin, from a kinetic series.
- Report total flux in a region, background subtracted.
- Keep the brightest pixel within the camera's range.